Cuff pops off, Cuff deflating too slowly – Welch Allyn Connex ProBP 3400 digital blood pressure device - User Manual User Manual

Page 71

Advertising
background image

Possible cause

Corrective action and explanation

Device is being used in inclimate environment

Ambient temperature is outside of operating

range. Retry measurement may display.
Operate the device within temperature range in the

Environmental specifications section.

Internal errors or messaging error occurred

NIBP feature not functional. Call for service

may display.
Contact a Welch Allyn Service Center.

Cuff pops off

Possible cause

Corrective action and explanation

Inappropriate cuff size

See “Blood pressure cuff selection” to select correct

cuff size. If cuff continues to pop off, notify biomedical

department or Welch Allyn Technical Support.

Cuff not applied securely

Smooth hook and loop securely before inflating cuff.

Cuff applied inside out

Re-apply cuff. Verify that the Welch Allyn label is

facing away from arm.

Cuff deflating too slowly

Possible cause

Corrective action and explanation

Patient movement

Have patient sit still.
Do not have arm tight against chest wall, as

respiration may affect speed and accuracy of blood

pressure measurement.

Arrhythmia

Check for regularity of heart rate (palpate pulse or

check device).
Moderate to severe heart rate irregularities may make

blood pressure difficult to measure accurately.

Small leak in pneumatic system

Check cuff tubing and pressure hose for leaks.

Cuff inflates beyond SureBP inflation pressure preset

while device is in use

Possible cause

Corrective action and explanation

Device switched from SureBP to Step algorithm

If the device is unable to determine a blood pressure

while the cuff is inflating due to patient movement,

excessive noise, or an arrhythmia, the device will use

Directions for use

Troubleshooting 65