Directions for Use

Chapter 3 Operation

15

Braun ThermoScan Pro 4000 Model

Figure 5. 9600 Plus Calibration Tester with the Braun ThermoScan
Pro 4000

To begin calibration verifcation of the thermometers:

1. Clean the probe tip with a cotton swab slightly moistened

with isopropyl alcohol, remove excess alcohol with a clean

Caution Store thermometers for testing in the same
room as the 9600 Plus Calibration Tester for
approximately 30 minutes prior to testing to allow for
thermal accommodation.