Directions for Use

Chapter 4 Maintenance

25

Specifications

Power
Requirements

Voltage: 100 - 240 V 50/60 Hz

Current

Heating: 480 mA

After Warm-Up: 185 mA

Power
Consumption

Heating: 2.4 W

After Warm-Up: 0.92 W

Calibration Tester

Length: 254 mm (10.00”)

Width: 105 mm (4.13”)

Height : 85 mm (3.34”)

Weight (without Power Pack): 0.75 kg (1.65 Lbs)

Power Pack
(approximate
dimensions)

Length: 86 mm (3.38”)

Width: 51 mm (2.00”)

Height (with plug inserted): 62 mm (2.44”)

Weight: 0.18 kg (0.40 lbs)

Environment

Storage Temperature: 0°C – 50°C (32°F – 122°F)

Ambient Operating Temperature: 18.3° C to 26.7°C (65° F to 80° F)

Operating Humidity: 20% to 95% RH. Non-Condensing

Operating
Specifications

Controlled Temperature of 9600 Plus

Low Setting: 36

o

C (96.8

o

F)

Intermediate Setting: 38.5

o

C (101.3

o

F)

High Setting: 41

o

C (105.8

o

F)

Temperature Accuracy: ±0.05° C (0.1

o

F)

Maximum Pressure/Altitude: < 2000 meters.

Heating and
Cooling Times
Tested at an ambient
of 23.9°C (75°F)

Room Temperature to Low Setting: Approximately 20 min.

Low Setting to Intermediate Setting : Approximately 6 min.

Intermediate Setting to High Setting: Approximately 6 min.

Cool from High Setting to Low Setting: Approximately 25 min.

Installation Category II
Pollution Degree II