Chapter 4 Maintenance

23

9600 Plus Calibration Tester Checks

If several thermometers do not pass the calibration verification
tests, check the 9600 Plus Calibration Tester to determine if it is
out of calibration. Before sending the unit to an authorized Welch
Allyn service center, complete the checks listed below.

1. Check the calibration date on the side of the 9600 Plus

Calibration Tester. If the 9600 Plus Calibration Tester was
calibrated over 12 months ago, do not use it. Send it to an
authorized Welch Allyn Service center immediately for
calibration.

2. Ensure no one has tampered with the 9600 Plus Calibration

Tester. If someone has tampered with the unit or the
calibration label(s), the calibration settings and warranty are
void. Send it to an authorized Welch Allyn Service center
immediately for calibration.

3. Ensure the 9600 Plus Calibration Tester is not exposed to

extreme temperatures that can damage the precision
thermistor. Exposure to high heat can cause the resistance
versus temperature characteristic of the unit to shift, causing
the 9600 Plus Calibration Tester to become inaccurate. Send
it to an authorized Welch Allyn Service center immediately for
calibration.

4. Verify the probe tip of the Braun ThermoScan Pro 3000 and

the Braun ThermoScan Pro 4000 are clean. Always use new
probe covers when calibration testing these thermometers.

E1.1

Device not calibrated

Return to Welch Allyn for service

E1.3

EEPROM failure

Return to Welch Allyn for service

Error

Problem

Solution