Changing the calibration set point temperature, Changing the – Welch Allyn 9600 Plus Calibration Tester - User Manual User Manual

Page 25

Advertising
background image

Directions for Use

Chapter 3 Operation

19

d. The LCD will flash “CAL” after the self check sequence.

The display will flash “CAL”, even if you have taken a
temperature.

e. Release both buttons.

3. Apply a new probe cover. Place the device firmly into the Ear

Device Port

(Figure 6)

.

4. Wait approximately three seconds, press the thermometer

Activation button, and listen for the beep.

5. Remove the Pro 3000 Thermometer from the 9600 Plus

Calibration Tester and read the temperature in the
thermometer’s display. If the temperatures are within ±0.2° C
(±0.4° F), the thermometer is within calibration.

6. Wait one full minute before taking another reading with the

same thermometer. Repeated measurements in short
sequence may cause higher readings.

7. The device will exit Cal mode after approximately 4 minutes.

8. Test all available thermometers for calibration verification at

the current Calibration Set Point Temperature. Proceed to the
next Calibration Set Point Temperature, see

"Changing the

Calibration Set Point Temperature" (page 19)

.

Changing the Calibration Set Point
Temperature

To scroll from one set point to the next, press and hold the
Temperature Selection button until a beep is heard. The newly
selected set point appears in the upper left corner of the LCD
display. The device’s current temperature is displayed, will start
to flash, and continue flashing until the cavity reaches the
equilibrium at the new set point.