Chapter 4 Maintenance

Welch Allyn 9600 Plus Calibration Tester

Problem Solving

Error

Problem

Solution

E

Calibration Tester cannot reach
Set Point Temperature because
room or device is too warm.

Ambient temperature and/or device
temperature should be within 18.3

°

C

to 26.7

°

C (65

°

F to 80

°

F). Allow 30

minutes for thermal accommodation
prior to testing.

E

Calibration Tester cannot reach
Set Point Temperature because
room or device is too cold.

Ambient temperature and/or device
temperature should be within 18.3

°

C

to 26.7°C (65°F to 80

°

F). Allow 30

minutes for thermal accommodation
prior to testing.

Voltage

Power adapter voltage problem Call Welch Allyn for replacement

E0.1

Broken Heater, not heating at
all

Return to Welch Allyn for service

E0.2

Heater run away (heater does
not turn off at appropriate time)

Return to Welch Allyn for service

E0.3

Open thermistor

Return to Welch Allyn for service

E0.4

Shorted thermistor

Return to Welch Allyn for service

E0.5

Open RCAL resistor

Return to Welch Allyn for service

E0.6

Shorted RCAL resistor

Return to Welch Allyn for service

E0.7

Open PTB resistor

Return to Welch Allyn for service

E0.8

Shorted PTB resistor

Return to Welch Allyn for service

E0.9

Catastrophic internal failure

Return to Welch Allyn for service