Administrator, Restore user defaults – Welch Allyn 1500 Patient Monitor v.1.3.X - User Manual User Manual

Page 112

Advertising
background image

108

Settings

Welch Allyn 1500 Patient Monitor

Restore user defaults

The saved user settings are enabled. The user-defined settings are saved in the Setup
System menu (see

“Saving the user-defined settings as default”

on page 21).

Administrator

The administrator and service screens provide system information and option settings.

The Administrator screens are entered from the settings menu:

Setup Menu > Administrator

Administrator sub-menus can only be accessed with a password. The passwords are
divided into clinical, service and factory passwords and only clinical settings are available
for the user. The other menu options on this menu are for Welch Allyn service personnel.
Options are enabled in the service menu; please contact Welch Allyn. The passwords for
the alarms and system sub-menus are as follows:

Setup > Administrator > Alarms: 49, 48, 46 (Clinical Password).

Setup> Administrator > System 49, 48, 46 (Clinical Password).

No password is needed to view the Configuration. Other sub-menus are intended for
service personnel only.

The Administrator menu is as follows:

Sub menu

Parameter

Description

Configuration

Monitor data such as serial number, software version etc. This
is provided for information only and no settings can be made.

Communications

This provides the communication settings for service
personnel only.

Alarms
(password
protected - see
above)

Alarm Silence Time

1

, 1.5 or 2 minutes. Time for which an audible alarm is silenced.

Alarm Suspend Time

1

, 1.5, or 2 minutes. Time for which all alarms are suspended.

Can disable HR / PR
Alarms

Enabled/

Disabled. The menu item can prevent users from

being able to turn off HR or PR alarms. When the setting is
"Disabled", a user cannot turn off the HR or PR alarms.

Alarm Delay

On/

Off. If the alarm validation is enabled, alarm limits must be

exceeded for at least 6 seconds for an alarm to be issued.

Second Speaker Time

0 to 3 minutes (2 minutes). Time after which the secondary
speaker is enabled. For lethal and high level alarms the
secondary speaker is activated in 30 seconds.