Respiration rate alarms and messages

Message

Possible cause

Suggested action

RR out of range (too
high)

The patient's RR is too high for
accurate measurement.

Check the patient.

Electrical interferences from
other devices.

Remove source of the interference.

Signal disturbed due to frequent
artifacts caused by bad
electrode contact.

Check and reapply/replace electrodes if
required.

RR lead off

Electrode loose/defective.

Check and reapply/replace electrodes.

RR artifact

Patient has moved.

Calm the patient.

Interferences by other devices.

Remove source of the interference.

Bad electrode.

Check/replace electrodes.

RR > [upper limit]
RR < [lower limit]

RR is higher or lower than alarm
limit.

Check the patient.