Approved catheters and injectate sensor type, Preparing for cardiac output measurement – Welch Allyn 1500 Patient Monitor v.1.3.X - User Manual User Manual
Page 83
Directions for use
Monitoring and Measurements
79
Approved catheters and injectate sensor type
Catheters
The monitor will work with any blood temperature sensor that is Edwards-compatible.
Injectate Sensor
The injectate sensor must be a Baxter/Edwards/Abbott "in-line" thermistor.
Preparing for cardiac output measurement
1.
Press the Setup button
, enter the menu Setup/Parameters and activate CO
enabled (yes).
2. Connect the catheter to the CO connector on the side panel.
3. Setup the catheter according to the instructions given by the manufacturer.
Note
The monitor does not support bath probe or arrow flow-through sensor.
Note
The CO is only displayed when the CO
option is enabled. If the CO measurement
field is not displayed, ensure that it is enabled in the parameter settings.