Cardiac output drug calculator – Welch Allyn 1500 Patient Monitor v.1.3.X - User Manual User Manual

Page 131

Advertising
background image

Directions for use

Technical data

127

Cardiac output

Drug calculator

Module

Schiller

Amplifier

Fully isolated, defibrillation protected >5kV

Measuring method

Thermodilution

Sampling frequency

250 Hz

Measuring method

Thermodilution

Parameters:

Cardiac output
Injectate temperature
Catheter temperature

Measuring range

Cardiac output: 0 to 20 l/min
Injectate temperature: 0° to 40°C / 32° to 104°F
Catheter temperature: 33° to 40°C / 91.4° to 104°F

Resolution

Cardiac output: 0.01 l/min
Injectate temperature: 0.01°C / 0.018°F
Catheter temperature: 0.002°C / 0.018°F

Accuracy

Cardiac output: ± 5% at 0°C / 32°F (injectate temperature)

Measurement

Start is detected by temperature difference >0.05ºC at distal thermistor

Calculations

Setting of a dose, rate, amount, and volume for which any 3 set will cause
the 4th to be calculated.
Concentration mg/ml
Dose and rate
Titration table (dose and rate)