CANNON

®

Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO

®

Instruction & Operation Manual

Version 2g — May, 2009; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

32

CAUTION

Make certain to enter the correct calibration standard viscosity (in
centistokes) BEFORE running the standards. Incorrect data entry may
skew future calibration results.

5. Click Run, then select the appropriate sample trays to test the

standards.

6. When testing is completed, use the VISCPRO

®

software to log on as

a Manager (Main/Log In) in order to access the calibration menu.

7. Click View Instrument from Main to select the desired CAV instru-

ment from the Configure menu; then select Calibration.

8. Use the spin controls

to select the desired Tube and Bulb for

calibration.

NOTE

Standard viscometer tubes have three timing bulbs (fast-run tubes have
two timing bulbs).