Viscosity action for viscosity standards, Copy & paste sample id data entry options – Cannon Instrument CAV-2000 User Manual
Page 67
61
CANNON
®
Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO
®
Instruction & Operation Manual
Version 2g — May, 2009; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
3. Enter necessary information in the Assign Sequential Sample IDs
window:
Prefix—Input a character string to be used as an identifier
Suffix—Input a character string to be used as an identifier
Starting ID—Input a number to start the numeric sequence
Number of Samples—Select the number of samples in the high-
lighted block which are to be sequentially numbered
Use as Format—If checked, this feature activates a “wrap-around”
function for sequential numbering like an odometer (if 001 was
input, the sequence would restart at 001 every 100 samples).
Assign Sequential Sample IDs window
4. When done, click OK.
NOTE
The VISCPRO
®
software “remembers” the last number in the last se-
quence and will place the next number in the sequence in the Starting ID
field the next time the Sequential Sample IDs option is selected.
Viscosity Action for viscosity standards
The Viscosity Action window for KV verification is identical to the
Viscosity Action window for measurement of kinematic viscosity (see
previous section), except that there is an additional field provided for
input of the known viscosity of the standard. This value must be entered
in the appropriate field in order that the resultant viscosity determination
may be used for verification and calibration
Copy & Paste Sample ID data entry options
The VISCPRO
®
Sample ID data entry screen supports Windows
®
Copy
and Paste options.
Copying Sample IDs
To copy a Sample ID entry, including the configuration information,
highlight the desired line(s) from the Sample List Box and right-click
your mouse to access the Sample Action window. Then select Copy All
Selected Samples from the menu options. OR you can use the C-C
key combination.