Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO
®
Instruction & Operation Manual
Version 2g — May, 2009; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
60
Viscosity Action window
The Viscosity Action window permits you to set test parameters for
individual samples, overriding the general settings. For information on
general settings, refer to Tray Settings: Test; beginning on page 43 of
this manual.
The Wash Options are only available for instruments configured for
dual-solvent operation. Click on the radio button to select desired solvent
option for individual samples.
NOTES
Parameters for the solvent wash are determined by Wash settings from
the Tray Settings: Wash window (see Tray Settings: Wash section, this
chapter, for more information). The Wash Options default setting is for a
solvent wash using both solvents.
To select more than one sample at a time from the list box, highlight the
first sample and the press the
S key and click on the last sample, or
just press and hold the
C key and click on individual samples to be
selected. When you have completed sample selection, right-click the
mouse on any of the selected samples to select the desired sample
action for all highlighted samples.
Sequential Sample IDs—creates a numeric sequence for the
highlighted sample ID(s) in the list box. You have the option of
selecting a standard alphanumeric prefix and/or suffix to accompany
the sequence numbers, and you also can select the starting numeral in
the sequence.
Procedure
1. To use the Sequential Sample IDs feature, highlight the desired
sample tray location(s) in the list box.
2. Right-click any of the highlighted locations and select Sequential
Sample IDs from the Sample Action popup.