Filling the bath

CAUTION

All fill and drain procedures
should be performed when
the bath fluid is at ambient lab
temperature. The procedure
below provides instructions
for maintaining bath fluid
levels. For instructions on
filling an empty bath, consult the CAV Installation & Setup Guide.

Fill procedure

1. Obtain the (supplied) funnel.

2. Locate the expansion vessel

(CAV-2100) or the top bath
flange (CAV-2200) at the top
of the temperature bath.

3. For the CAV-2100, open the

flip-cap on the expansion
vessel (see photo). For the
CAV-2200, loosen the
knurled thmbscrew (see
photo) securing the alumi-
num hole cover to the top
bath flange and slide the
cover free of the opening.
Then insert the appropriate
funnel in the opening.

4. Using the funnel, introduce the appropriate grade of silicone bath

fluid into the expansion vessel (CAV-2100) or bath (CAV-2200).

5. Continue adding fluid until the liquid level rises to the cold fill level.

For the CAV-2100, that level is even with the bottom of the top bath
flange. For the CAV-2200 that level is about 2” or 5 cm below the top
flange, and is marked by a decal fill line.

6. When done, remove the funnel and replace the flip-cap or hole cover.

Check the bath fittings for leaks to complete the fill procedure.
Tighten fittings if necessary.

CAV-2200 fill hole cover

Expansion vessel with liquid level
elevated at high temperature

meniscus

level