Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO

®

Instruction & Operation Manual

Version 2g — May, 2009; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CAUTION

Advanced settings will ordinarily be configured by a technician from
CANNON

®

Instrument Company when the instrument is initially installed.

Advanced settings should only be adjusted by or in consultation with a
representative from CANNON

®

.

Maximum bulb 1 fill time(s):

Adjust this setting to select the desired maximum fill time for bulb one.
Acceptable values are between 20 and 400 seconds.

Min. bulb 1 selection time(s):

Adjust this setting to select the desired minimum time permitted for
selection of bulb one for sample testing. Acceptable values are between
3.5 and 20 seconds.

Minimum bulb 2 fill time(s):

Adjust this setting to select the minimum fill time permitted for bulb two.
Acceptable values are between 0 and 4.0 seconds.

Maximum bulb empty time(s): Adjust this setting to select the maximum drop time permitted for the test

before the software will abort the sample run. Acceptable values are
between 90 and 600 seconds.

Maximum tube empty time(s): Adjust this setting to select the sample recovery time (the amount of time

between the conclusion of the sample run and the initiation of the wash
cycle). Acceptable values are between 5 and 125 seconds.

When you have selected the desired tray settings, save the current
instrument settings by clicking OK. To exit the configuration screen
without saving your changes, click Cancel. To restore software defaults,
click Defaults.

Bulb overdraw

Use the Bulb Overdraw button/window to set the the amount of time in
seconds that the instrument will draw sample AFTER the sample menis-
cus has cleared the top sensor.