2.8.4 Imbalance detection

The device switches off if there is excessive imbalance. (See

Ä

Chapter 10 ‘Repairs’ on page 55)

2.8.5 Operation on GFCI (Ground-Fault Circuit Interrupters)

It is possible that the leakage currents will become marginal during oper-

ation. This may trigger the earth leakage circuit breaker. These values

can quickly be reached when all devices are added to the circuit.
Solution: Circuit without an earth leakage circuit breaker or, if possible,

increase the earth leakage circuit breaker threshold.