The mains voltage has been set at the factory to that of the

specific country. The mains voltage only has to be adjusted

if it deviates from the value on the type plate. If adjustment

is necessary, proceed as in Ä Chapter 4.5.1.1 ‘Adjusting the

mains voltage with the rotary switch (8)’ on page 24 and

Ä

Chapter 4.5.1.2 ‘Adjusting the mains voltage in setup

mode’ on page 25.

NOTICE!
Fritsch mills are speed controlled. The devices are equipped

for this with frequency converters. In order to comply with

the EMC directive, many measures must be taken to pre-

vent operational transient emissions.
The possible leakage currents resulting from filtering meas-

ures can trigger a conventional residual current circuit

breaker in the mains line. This is no defect!
To prevent this, special residual current circuit breakers,

which are adapted for operation with frequency converters,

are commercially available.
Operation without a residual current switch is possible, but

must be done in accordance with the relevant regulations.

4.5.1 Adjusting the mains voltage

4.5.1.1

Adjusting the mains voltage with the rotary switch (8)

CAUTION!
Only qualified personnel may change the voltage range on

the device!

CAUTION!
The voltage range may only be adjusted after the mains has

been disconnected.