6 electrical connection – FRITSCH PULVERISETTE 16 cast iron User Manual

Page 15

Advertising
background image

Page 14

PULVERISETTE 16

3.6 Electrical

connection

The electrical connection may only be created by a qualified electrician.
The voltage and frequency of the cross beater mill can be found on the type
plate.
Make certain that the values match the existing power network.

Connect the cross beater mill to the power network using the supplied con-
nector cable.
The supplied power cable does not have a plug, since the type of plug de-
pends on the setup location and the respective national regulations.
External fuse protection must be provided during connection of the power
cable to the mains in accordance with the regulations of the setup location.

Direction of rotation of the drive motor

Viewed from the drive side, the 3-phase AC motor should rotate anticlock-
wise
. (Viewed from the front, looking at the cross beater mill, anticlockwise;
viewed from the rear, clockwise.)
The single-phase AC motor is factory-set to rotate anticlockwise.
See:
• DIN VDE 0530 part 8 “Terminal marking and direction of rotation”
• DIN VDE 0530 Teil 7 / EN 60 934-7 “Type codes”.
The direction of rotation of the 3-phase motor can be changes by swaping
two of the leads, L1, L2 and L3 (either at the motor terminals or in the plug).
The direction of rotation may only be changed by a trained electrician.

DANGER

Power Supply Voltage

The connector cable may only be altered by a trained expert.

CAUTION

Non-observance of the values on the type plate may dam-
age electrical as well as mechanical components.

CAUTION

Electrical Safeguard!

Danger of damages due to short-circuit.

• Ensure that the socket is connected to a power line se-

cured with a circuit breaker.

Advertising