Flags 10 -16, Input 1 61…66, Ktare / wtare 112 / 113 – Industrial Data Systems IDS Scale Basic User Manual

Page 37: Minus 102, Motion 107

Advertising
background image

Scale Basic 4.2E

Reference

Flags 10 -16

Purpose: Enables control of status LEDs on display panel (Gross, Net, Motion. Zero, Lb, Kg)
Remarks: Flag 10 is used to allow turning on and off the status LED displays, flags 11-16.
Example: Use the F1 & F2 keys to toggle on and off the all the displays. Individual LEDs could
be used to show an over under or between status, high or low status as a bar graph or
to emulate a masters display in a master/ slave network.

Set CFG 57 to 1. F1 key will execute user Fn 1, turn on all LEDs.

Set CFG 58 to 2. F2 key will execute user Fn 2, turn off all LEDs.


Fn. 1:

Flag on, 10

Enable LED control flag 10

Flag on, 11

Flag on, 11 = Gross LED on

Flag on, 12

Flag on, 12 = Net LED on

Flag on, 13

Flag on, 13 = Motion LED on

Flag on, 14

Flag on, 14 = Zero LED on

Flag on, 15

Flag on, 15 = Lb LED on

Flag on, 16

Flag on, 16 = Kg LED on

Display, Gross Display the gross weight

Fn. 2: All off

Turns all flags to false and off.

NOTE: The all off instruction turns off all flags, setpoints & timers, one may use
the ( flag off, 10 thru 16 for individual control of each LED ).

Input

1..6

61…66

Purpose: true if TTL input is idle (TTL high)
Remarks: the TTL inputs are normally high. They are often used to detect a switch closure

which connects the TTL input to Ground (TTL low). The TTL inputs use negative logic!

Example:

If, Input5

test for input 5 idle

If not, Input5

test for input 5 ON

Ktare

/

Wtare

112

/

113

Purpose: true if the tare register data was entered from the keyboard (Ktare) or read from the

scale (Wtare).

Minus 102

Purpose: true if previous calculation result was negative.

Motion

107

Purpose: true if the scale is in motion.