Calibration, Cal 0, 1. calibration using a weight – A&D Weighing FG-150KBM User Manual

Page 14

Advertising
background image

12

8. CALIBRATION

This function adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.

? When the scale is first used.
? When the scale has bee moved.
? When the ambient environment has

changed.

? For regular calibration.

Loose the lock screws on the rear side of
the display pod, and remove the CAL
switch cover. Then, there is a calibration
switch on the board inside.

? For the FG-KB series (without column), you may once remove the weighing pan to

access lock screws easily.


? Do not use a ballpoint pen and so on to press the calibration switch.

That may short-circuit and damage the scale.


8-1. Calibration using a weight

1. Warm up the scale for at least half an hour with nothing on the weighing pan.

? Change Function setting “F1” or place something on the pan to disable the auto

power-off function.

2. Press and hold the calibration (CAL) switch until

Cal 0 appears, and release the switch.

? The weighing unit must be “kg” or “lb” to enter

calibration mode.

3. Make sure that there is nothing on the weighing pan,

and wait until the STABLE indicator turns on.

4. Press the MODE switch. The scale calibrate the

zero point, and the display shows “5pn 1” and the
weight value to calibrate (SPAN calibration).

? The weight value is equal to the capacity. When

you enter with “kg” mode, then the value is “kg”.
Entering with “lb”, then “lb”.

? If you do not need SPAN calibration, turn the power off to exit from the calibration

procedure.

5. To calibrate with the different weight, change the displayed value using the following

switches.

RE-ZERO or ZERO To select the digit blinking to change.

PRINT

Increment the value of the selected digit.

5pn 1

60.00

Calibration (CAL) switch

Connector

Lock screw

CAL switch cover

Cal 0