Comparator – A&D Weighing FG-150KBM User Manual

Page 13

Advertising
background image

11

7. COMPARATOR

The results of the comparison are indicated by HI, OK or LO on the display.
The formula to compare is as follows:

LO < Lower limit value = OK = Upper limit value < HI

Operating conditions (see the “Function list F6”):

F6-0: No comparison (comparator function disabled).
F6-1: To compare all data.
F6-2: To compare more than +4d or less than -4d.
F6-3: To compare all stable data.
F6-4: To compare stable data more than +4d or less than -4d.
F6-5: To compare stable data more than +4d.
d = minimum weight display (see “12-1 Specifications”)
In case of counting mode, “d” is equal to minimum weight display of kg mode.

? The upper limit and lower limit numerical values are common to each of the

weighing and counting mode.

? Ignore the decimal point of setting value to apply it to each mode.

Example of FG30K / setting value is “001000”.

Display mode

Limit value

Display capacity

NORMAL resolution kg

10.00 kg

30.00 kg x 0.01 kg

HIGH resolution kg

1.000 kg

30.000 kg x 0.005 kg

HIGHER resolution kg

1.000 kg

30.000 kg x 0.002 kg

NORMAL resolution lb

10.00 lb

60.00 kg x 0.02 lb

HIGH resolution lb

10.00 lb

60.00 kg x 0.01 lb

HIGHER resolution lb

1.000 lb

60.000 kg x 0.005 lb

Counting mode

1000 pcs

Entering the upper and lower limit values

1. Press the MODE switch to select weighing

unit ”kg”, ”lb” or ”oz”

2. Press and hold the MODE switch to enter the

upper limit setting mode.

3. Enter an upper limit value using the following switches.

RE-ZERO or ZERO To select the digit blinking to change.

PRINT

Increment the value of the selected digit.

The minus sign can be set at the next digit of the least
significant digit. The PRINT switch alternates the
minus sign on and off. The blinking “-” shows minus
and no sign is plus.

4. After setting all of digits, press the MODE switch. Then, the upper limit is stored

and the display goes to the lower setting mode.

5. Set the lower limit in a similar way, and press the MODE switch to return to

weighing mode.

? The upper and lower limits are maintained even if the scale is powered off.

0.00

0000.00

HI

-0000.00

HI