6. Place the calibration weight on the pan with the same

value as displayed, and wait until the STABLE
indicator turns on.

7. Press the MODE switch. The scale calibrate SPAN

and end will appear.

Remove the weight from the pan, and turns the power
off.


? Note

The value set in step 5 is cleared after the power is switched off.

If the scale will suppose to move to another location, set the gravity
acceleration value for the current location and calibrate the scale
according to the procedure above. See the next section to set the
value.


8-2. Gravity acceleration correction

When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value of the scale to the value
of area where the scale will be used. Refer to the gravity acceleration map appended
to the end of this manual.

? Note

It is not necessary to set the gravity acceleration correction when
calibrating the scale with a calibration weight at the place where it is
to be used.


1. Press and hold the calibration (CAL) switch until

Cal 0 appears, and release the switch.

? The weighing unit must be “kg” or “lb” to enter

calibration mode.

2. Press the PRINT switch.

The display shows the gravity acceleration value
memorized in the scale.

3. To change the displayed value using the following switches.

RE-ZERO or ZERO To select the digit blinking to change.

PRINT

Increment the value of the selected digit.

4. Press the MODE switch. The display returns to

Cal 0 .

5. If necessary to calibrate the scale using a calibration weight, go to step 3 of “8-1.

Calibration using a weight”. To finish the setting, turn the power off.



9.798

end

888

40.00

Cal 0