SEA Series Steam Sterilizers

Operation manual

8

5
.2 Start a cycle

When water indication light is off, the machine can be started.

5.2.1Press SELECT to choose one of the programmes.
The key toggles among the cycles at 121 ,134 and B

+

mode. One of the indication lights turns on to show the
programme you selected.(Fig 5-2)

5.2.2 You may need to select instruments package mode. According to the instrument you put
into the chamber whether it is bareness or sop up, press package to select package mode.

5.2.3 You may select DRY key when starts the fast sterilization cycle. Press DRY
key, it will reduce the sterilization time. If you press the butoon"DRY" on the panel, there
are three selections, 1min(with 1min LED on), 5min(with 5min LED on) , and 10min(with
two LEDs off).

5.2.4 After selecting above steps, put the instruments
into the Steam Sterilizer which need sterilize.

Caution: Instruments that are ready to sterilize should
be placed in the tray not touching one another allowing
space for steam to pass through. It is suggest that the
tray holder be used to put the tray with instruments
inside the chamber to avoid burning injuries.

Fig 5-3

Fig 5-2