Appendix, Form1 cycle mode chart, Form2 work mode chart – LABEC AA-12S User Manual

Page 27: Sterilize mode, as following form(8l/12l), Operation manual

Advertising
background image

SEA Series Steam Sterilizers

Operation manual

2 3

Appendix.

Form1 Cycle mode Chart

Item

Digital display

Mode

Description

1

Ld

Pre-heating

Preparation and conditioning of the

autoclave

2

HE

Heating

Chamber is being heated and pressurized

3

Cycle time

Sterilizing

Counting down of sterilizing

4

Dr_

Venting

Chamber starts to vent and dry

5

Drying time

Drying

Instruments are being dried.

6

Ed_

Complete

Cycle is complete and chamber door can

be opened

7

Faults & Error codes

Error codes

Displays error codes, resets and alarm

8

LE_

Balance state

When change the state, the pressure of inside

and outside the Autoclave should be the same.

Form2 Work mode chart

Sterilize mode, as following form(8L/12L)

o

134 C
package

o

134 C

non package

o

121 C

package

o

121 C

non package

134B+

BD test

Vacuum
test cycle

CLEAN test

Sterilizing

temperature

134

134

121

121

134

134

Sterilizing pressure

2.10

2.10

1.10

1.10

2.10

2.10

Classes

S

S

S

S

S

B

Pre-heating min

0-7

0-7

0-7

0-7

0-7

0-7

Vaccumize time min

4

4

4

4

4

10

4

Clean 20

Admission time

min

5

5

5

5

5

5

Drying

Sterilizing time

min

4

6

16

20

18

3.5

Keep
pressure 10

Drying time

min

10

10

10

10

10

1

10

Quick drying time

1/5

1/5

1/5

1/5

1/5

Total time

min

30(21/25)

38(29/33)

42(33/37)

52(43/47)

50(41/45)

26.5

14

30