8.1 Enter function setup mode:Press STEP key for 20s when on non-work state

time dispaly Ld

pressure dispaly 1

enter function setup mode.

8.2 Function mode setup With 7options, press SELECT key, you can select 1-7-1cycle
dispaly. The definition and display of 7 options as following

Option 1 year dispaly yer

press STEP or

key increase press DRY

or

Key decrease setup numerical value

Option 2 month dispaly nth

press STEP or

key increase press DRY

or

Key decrease setup numerical value

Option 3 day display day

press

STEP or

key increase press DRY

or

Key decrease setup numerical value

option 4 hour display hor

press STEP or

key increase press DRY

or

Key decrease setup numerical value

Option 5 minute display nin

press STEP or

key increase press DRY

or

Key decrease setup numerical value

Option 6 Printing,it display Pvt ,press STEP key then you can choose the cycle
b e t w e e n o n a n d o f . o n s t a n d f o r p r i n t , o f s t a n d f o r n o p r i n t i n g .

Option 7 Exit,,when Eit displays,please press START/STOP to save and exit setup
mode.

Option 8 sterilization time adjust,display S-T when you setup this option,the sterilization
time(SP)win increase at former value.The adjustable range are following 3 kinds.One is
inereased by 1 minute,Second is inereased by 2 minutes.Third is incereased by 3 minutes.

Option 9: temperature data adjust automatically option, display D-P ,When you setup this
option,you can adjust the temperature more or less than 2C-3C inorder to meet your request.
But we suggest user should setup under guidance of teahnician or not setup.

Press SELECT key after setup,you will come back to option 1.At this time,you must press
STEP key 3 seconds in option7.Then you can start the setup of option8 and option.Otherwise
it only can start cycle between option1and option7.

Caution:If some function setup was changed by pressing the key without intention,please
switch off the power,then restart it.It will exit from the wrong setup.In this case,the function
won't be affected by this.