SEA Series Steam Sterilizers

Operation manual

9

Fig 5-4

5.2.5 After the instruments are loaded, you may close and lock
the chamber door by turning the door knob clockwise to its
locking position. The ready indication light goes on. (Fig 5-
4)

Caution: The door knob must be turned to its lock position,
otherwise, there might be steam leak during a cycle and the
alarm goes on and displays error code E6 . If the chamber
is in keep warm mode, it is sometimes difficult to close the
chamber door because there is still steam inside. You can keep
the door open for a while until steam and moisture is vaporized,
and the door is easy to close. You may also push the door inwards
with some force while turning the door knob to its locking position. If there is still steam leak
after you did the above, you may try to read just the door mechanism by a spanner as per item
6. Door adjustments.

5.2.6 Press START/STOP key, the ready light goes off
and display HE the machine starts a sterilization cycle. It takes
about 30-60 minutes to complete an automatic cycle that includes
heating up, sterilizing and drying. The cycle time depends on the
volume of the instruments loaded, the chamber
temperature at the beginning of the cycle and the cycle program
you selected.

5.2.7 When a cycle completes, the display Ed the indication light

RUNNING Turns off and the indication light READY turns on
with sounds. At this time you may then open the chamber door and unload

the instruments which are sterilized. (Fig 5-6) At mean time, temperature
indicate prt , it means that it is performing print, whether you have install
printer, the process can not be omitted. But you needn't care about it. It will
not effect the process of opening the door and unloading the instruments.
The indication will disappear when printing is over.