SEA Series Steam Sterilizers

Operation manual

1 0

Attention: To ensure autoclave's on the safe side, When water level low alarm sounds, Please
be advised add some water first. When you add the water, please don't forget to let condenser
water out.

Caution: Always use the tray holder to load or unload the tray with instruments in it, in order

to avoid burning injuries.

If you didn't press STEP key during a cycle and the indication light is off, it
means that the machine is keep warm mode.The sterilization Chamber is keeping
warm. This will enable faster cycle that follows.

If you press STEP key before the end of sterilization, STEP indication light goes on,
it means that the machine is last time cycle. The chamber starts to cool down completely before
the chamber door is open. It should normally be used during the very last cycle of the day.

5.2.8 If you decide to stop using the autoclave for a while, please turn the main power switch
off, or unplug the power cord should you not want to use the machine for a long time.

Caution: Never try to open the chamber door if and when the pressure display doesn't show

0.0 . It is suggested that chemical indicator or bio-indicator be used for each and every
cycle to ensure the reliability of each sterilization cycle run.