8

3 The screen will display program parameters and picture as follows after the door is

opened




3.2.3

Vacuum Test

1 Press“▲” key or “▼” key until screen display 6# vacuum test program shows as

following,close the door and press “on/off” key and the vacuum test program will

start.

2 The buzzer alarming will remind user that the vacuum test program is finished, and

screen will display the following:

3

Leakage rate must be lower than 0.13KPa/min。

3.2.4

B&D Test

1 Press“▲”key or“▼” key until screen display 5#B&D test program picture shows as

following:

1#i ns t r ument door : open
St er i l i zat i on t emper at ur e :134. 0℃
St er i l i zat i on t i me: 4. 0 mi nut es
Dr ynes s t i me: 2. 0 mi nut es

6#vacuum

door : open


Test leakage rate

6#vacuum t es t 20:00

Leakage r at e:

0. ***kPa/ Mi n

Pl eas e open t he door

5#BDt es t

door : open

St er i l i zat i on t emper at ur e :134. 0℃
St er i l i zat i on t i me: 3. 5 mi nut es
Dr ynes s t i me: 2. 0 mi nut es