Put B&D test bag into autoclave and close the door, press “on/off ” key and program

will start.

2 Buzzer will sound after B&D test is finished.

3 Remove the tray and observe the B&D test bag.

!!

Caution

Please use professional tongs and gloves with safety eyewear to remove the

tray in order to avoid injury.

Chapter 4 Preparation before Using

4.1 Preparation before Instrument Sterilization

The instruments must be cleaned before placing them into the chamber.

For example:Blood or other impurities.

Please refer to following methods:

1 Wash instruments immediately after use to dispose of any residue. It is

recommended that instruments should be washed with an ultrasonic cleaner, using

detergent solutions and distilled water.

2 After cleaning, rinse instruments again in order to ensure the instruments are free of

any residue.

3 Pay attention when arranging instruments on the tray:

Different instruments should be separated into different baskets, for example:

stainless steel instruments in one basket and carbon steel instruments are kept in another
basket. There must be a distance between the instruments. Sterile paper should be used
at the bottom of the basket if carbon steel instruments are to be treated, in order to prevent
the carbon steel from coming into direct contact with the stainless steel basket.

4 Place a sterilization indicator in each basket.