LABEC Shinva-45N User Manual

Page 17

Advertising
background image

15

Symptom

Possible Cause

Correction

11.2 no printing ink

11.2 change new printing ink

7.2 How to Remove the Cover of Autoclave

!!!

Make sure that the electrical cord is disconnected from the power supply and that there is no

pressure in the chamber.

Removal process:

1

Remove the screws on the side and back

2

Lift up the cover

7.3 How to Check the Safety Valve

In order to prevent the safety valve from blockage, every month under ordinary usage, it is necessary to allow the
steam pressure to escape through it.

1 Run 1# program

2 When pressure of chamber is 210Kpa, pull the ring of safety valve and open it 2 seconds

7.4 How to Replace the Safety Valve

Attention: This method should be used only by professionals.