LABEC Shinva-45N User Manual

Page 16

Advertising
background image

14

Symptom

Possible Cause

Correction


5 4 screen display:



4.1 too much water in waste water

tank

4.1 empty water from waste

water tank

6 5 screen display :



5.1 safety valve does not open

5.1 open safety valve for 2

seconds and close it in B&D test
process, if will not open please
fit a new one.

7 6 screen display :


6.1 short circuit in heating
controller

6.1 professional to inspect
circuit or fit a new controller

8 7 pressure not dropping

or drops slowly

7.1 high temperature in chamber
7.2 door gasket not sealing

correctly
7.3 air filter is blocked
7.4 steam exhaust pipe or water

drain pipe is blocked

7.1 open the door and try again
after half a hour

7.2 check door seal, may need a
new door gasket
7.3 detach filter and wash it

7.4 clean those pipes

9 8 in heating process,

pressure or temperature
not rising or rises slowly

8.1 controller of heater has short
circuit or is damaged

8.2 connection or safety valve
leakage

8.1 inspect and change damaged
parts

8.2 inspect and tighten
connector or safety valve

10 9 in pressure release

process, pressure or
temperature not

dropping or drops
slowly

9.1 filter of water drain is blocked 9.1 clean filter

11 10 in pressure balance

process, pressure not
rising or rises slowly

10.1 air filter blocked

10.1 fit a new air filter

12 11 data will not print

11.1 paper jammed or no paper

11.1 clear or fit new paper

War ni ng!

Ful l of was t e

wat er t ank

XXXkPa XXX. X℃

请排水

Warning !

Higher pressure

Pressure : 0KPa

Warning !

Higher temperature

Pressure : 0KPa


Advertising