2 debugging 3.2.1 – LABEC Shinva-45N User Manual

Page 9

Advertising
background image

7

3.2 Debugging

3.2.1

Original Parameters Settings

TMQ.CV3250 table –top autoclave should display original parameters at first use. The

setting method is as follows:

1 Switch on power switch, and press “set”, the system entry picture is as follows:

Remark:P0-local atmospheric pressure, T0-local boiling point temperature

2 Set atmospheric pressure,original value is 101kPa,press“▲” key and “▼” key to

change the atmospheric pressure.

3 Press “Esc” key after set value completion.and system will return to original menu.

3.2.2

Opening the Door.

1 Quit original menu, the autoclave will self-test: balance pressure of inside and

outside. Close the door lock, LCD displays the following. This process need

several minutes

2 The screen display picture is as follows after the door lock is turned on, user can

open the door.

!!

Caution:

Please don’t open the door using force until it displays “Please open door”.

Original parameters

Atmospheric pressure:101kPa
Local boiling point:100℃

version:090000

Reminding:

Balancing pressure

XXXkPa XXX. X℃

Please wait。。。

Reminding:

Door lock open

XXXkPa XXX. X℃

Please open the door