2 debugging 3.2.1 – LABEC Shinva-45N User Manual

Page 9

Advertising
background image

7

3.2 Debugging

3.2.1

Original Parameters Settings

TMQ.CV3250 table –top autoclave should display original parameters at first use. The

setting method is as follows:

1 Switch on power switch, and press “set”, the system entry picture is as follows:

Remark:P0-local atmospheric pressure, T0-local boiling point temperature

2 Set atmospheric pressure,original value is 101kPa,press“▲” key and “▼” key to

change the atmospheric pressure.

3 Press “Esc” key after set value completion.and system will return to original menu.

3.2.2

Opening the Door.

1 Quit original menu, the autoclave will self-test: balance pressure of inside and

outside. Close the door lock, LCD displays the following. This process need

several minutes

2 The screen display picture is as follows after the door lock is turned on, user can

open the door.

!!

Caution

Please don’t open the door using force until it displays “Please open door”.

Original parameters

Atmospheric pressure:101kPa
Local boiling point:100℃

version:090000

Reminding:

Balancing pressure

XXXkPa XXX. X℃

Please wait。。。

Reminding:

Door lock open

XXXkPa XXX. X℃

Please open the door