2 basic features, 3 running: environmental, Requirements – LABEC Shinva-45N User Manual

Page 6: 4 safety devices

Advertising
background image

4

高 水 位

H I G H L E V E L

L O W L E V E L

低 水 位

w a te r l e v el

2.2

Basic Features

u

LCD display, touch key pad for easy operation.

u

Independent steam generator device for heating quickly.

u

Continuous pulse vacuum system together with air exhaust, thoroughly ensures

saturation of steam and efficient sterilization.

u

The pulse vacuum system ensures dry products after sterilization.

2.3 Running: Environmental

Requirements

1 Room temperature: 5 —40 。

2 Maximum relative humidity: 85%

3 The voltage fluctuation is ±10%

2.4 Safety Devices

u

Auto protection device against over temperature:This system cuts off power

automatically if the designed temperature is surpassed

u

Door safety interlock:The door cannot be opened until there is no pressure in the

chamber and power for heating cannot be switched on until the door is firmly

closed.

u

The system will cut off the power when water not enough in steam generator.

Advertising