Set concentration of standards, Display standard curve – LABEC V-18 User Manual

Page 14

Advertising
background image

16/18 Manual

11


ENTER

RETURN

Fig. 14

Input the number of your standard samples by numeric keypad. After

pressing

【ENTER】

,

it returns to the setting interface.

Note

:

the number range is 1-9

,

other number is invalid.

Set Concentration of Standards

Before this step, you must blank first. Pull the blank cuvette in the light

path in the pre-testing interface, press

【ZERO】

.

Move the cursor in “

Concentration

”, following press

【ENTER】

.(Fig. 15)


ENTER



RETURN

Fig.15

Pull sample 1 in the light path, input the concentration of sample 1 by

numeric keypads, then press

【ENTER】

. It goes into the next setting

interface of sample 2, the operation is the same as sample 1. Repeat the
operation one by one till the last sample finished, and it goes back to the
setting interface.


Note:

①

The Abs. or T% of the samples will not display on the screen,

they were saved in the RAM

.

The range or concentration is 0-9999

,

other number is invalid.

③

The sequence of concentration is from low to high.

Display Standard Curve

Move the cursor on

“Display Curve”

by pressing

【 】

and

【 】

, followed

by pressing

【ENTER】

, the standard curve you just established will display

on the screen. The equation displays on the bottom of the screen. (Fig. 16),
and it will also be saved in the system.


√

√

√

√Standard Curve

○

○

○

○Unit
●

●

●

●Number
○

○

○

○Concentration

Standard Curve
N=2
Please Input New
:

:

:

:

N=

Pull Sample 1
C=1.500
Please Input New
C=

√

√

√

√Standard Curve

○

○

○

○Unit
○

○

○

○Number
●

●

●

●Concentration