Set test mode, Set upper limit – LABEC V-18 User Manual

Page 21

Advertising
background image

16/18 Manual

18

ENTER

,

it will return to the kinetics setting interface.

Note:

1. The range of total test time is 1-120,000, other value is invalid.
2.The maximum value of total time is correlative to the interval time,
because system only permits 1000 dots be adopted at most. For
example:
if the interval time is 1s, then the maximum total test time is:

1 X 1000=1000s;

if the interval time is 20s, the maximum total test time is:
20 X 1000=20,000s;
if the interval time is 200s, the maximum total test time is

NOT

:

200 X 1000=200,000s
Because the range of total time is 1-120,000, when the interval time is
200s, the maximum test time is 120,000!

3.

If you don’t want to change the value, pressRETURNdirectly.

Set Test Mode

Move the cursor on “

Test Mode”

by pressing

【∧

∧】

and

【∨

∨】

in setting

interface, followed by

ENTER

.(Fig. 34)

ENTER

RETURN

Fig. 34

Move the cursor on the test mode

by pressing

【∧

∧】

and

【∨

∨】

, then press

ENTER

to select. Press

RETURN

to go back to the kinetics setting

interface.

Set Upper Limit

Move the cursor on “

Upper Limit”

by pressing

【∧

∧】

and

【∨

∨】

in setting

interface, followed by

ENTER

.(Fig. 35)

ENTER

RETURN

Fig. 35

I

nput the new upper value by pressing the numeric keypad followed

by

ENTER

,

it will return to the kinetics setting interface.

Kinetics

Time Interval

Total Time

Test Mode

Kinetics

Abs.

T%

Energy

Kinetics

Upper Limit

Lower Limit

Display Data

Kinetics: Abs.

Upper=3.000

Please Input New

Upper=

Advertising
This manual is related to the following products: