Basic operation – LABEC V-18 User Manual

Page 9

Advertising
background image

16/18 Manual

6

Turn on spectrophotometer by pressing the Power Switch (IO). The instrument starts to

initiate and the steps are as below:

1.The instrument will position filter first, then initialize lamp position and AD converter,

then D2 / W lamps warm up, then Wavelength reset and get dark current, After initializing

printer, checking battery and setting up system, the screen display the main menu. And

the self-testing process display as fig 3.

2 .If the data in memory has been lost, the instrument will directly calibrate system

without any choice for you.

Note:

DO NOT OPEN SAMPLE COMPARTMENT LID DURING SELF-TESTING.
















Fig 3

Basic operation

Blank

Push the blank cuvette into the light path.

Press the key

【ZERO】

for blanking

Note:1

、

、

、

、

If the reference solution is too thick, the energy will be low.

2

、

.Blanking is automatic after a wavelength change.

DO NOT OPEN SAMPLE COMPARTMENT LID DURING BLANKING.

Set Wavelength

(Example: set wavelength in “Basic mode”)

Press

【GOTOλ

λ

λ

λ】

(Fig 4).

GOTOλ

λ

λ

λ


RETURN

√

Self-testing

●

●

●

●Filter √
○

○

○

○Lamp √
○

○

○

○AD √

√

Self-testing

○

○

○

○D2 Lamp √
○

○

○

○W Lamp √
○

○

○

○Reset WL √

●

●

●

●Basic Mode
○

○

○

○Quantitative
○

○

○

○Kinetics 08:00
○

○

○

○Utility 01/01

√

Self-testing

○

○

○

○Dark Current √
○

○

○

○Printer √
○

○

○

○Battery √

√

Self-testing

○

○

○

○System √

Main Menu

Basic Mode:

:

:

:

WL=546.0nm

Please Input New
WL=

Basic Mode:

:

:

:

0.000

Abs

546.0 nm