Set lower limit, Sample test – LABEC V-18 User Manual

Page 22

Advertising
background image

16/18 Manual

19

If you don’t want to change the value being set last time, press

ENTER

or

RETURN

directly .

Note: the range or A is -0.3-3, the range or T is 0-200, other value is invalid.

Set Lower Limit

Move the cursor on “

Lower Limit”

by pressing

【∧

∧】

and

【∨

∨】

in setting

interface, followed by

ENTER

(Fig.

36)

ENTER

RETURN

Fig. 36

I

nput the new lower value by pressing the numeric keypad followed

by

ENTER

,

it will return to the kinetics setting interface.

If you don’t want to change the value being set last time, press

ENTER

or

RETURN

directly .

Note: the range or A is -0.3-3, the range or T is 0-200, other value is invalid.

Sample Test

After you have finished setting “Time Interval”, “Total Test Time”, “Test
Mode”, “Upper Limit” and “Lower Limit”, press

RETURN

to go back to

kinetics pre-testing interface. (The left interface of Fig. 37)
Pull the blank cuvette in the light path, press

ZERO

for blanking.

Pull the sample cuvette in the light path, press

START

to go into testing

interface.
Press

START

again to began your test. (Fig. 37)

START

RETURN

Fig. 37

When the testing begins, the curve will be drawn on the screen in real

time. And the time will also be displayed in the middle of the bottom screen in
real time.

Press

STOP

can interrupt the test. When

START

being pressed again,

the test will start again from the very beginning.

Note: the curve can not be printed, if you want to check the data of

every dot, you can choose “Display Data” in the parameter setting

Kinetics

100.0 T%

546.0 nm

Kinetics 3.000

-0.500

0S 60S

546.0nm 0.002A

Kinetics

Upper Limit

Lower Limit

Display Data

Kinetics: Abs.

Lower=-0.500

Please Input New

Lower=