4 middle trigger mode, 5 acquisition with re-triggering, Middle trigger mode – ADLINK PCIe-9852 User Manual

Page 30: Acquisition with re-triggering, Figure 3-8, Middle trigger mode acquisition

Advertising
background image

20

Operations

3.4.4

Middle Trigger Mode

Collects data before and after the trigger event, with the amount to
be collected set individually (M and N samples), as follows

Figure 3-8: Middle Trigger Mode Acquisition

3.4.5

Acquisition with Re-Triggering

A digitizer acquires a trace of N samples/channel for a single

acquisition. Re-Trigger mode can also be set to automatically
acquire R traces, containing N*R samples/channel of data, without
additional software intervention.

The Re-Trigger setting can be used for Post-Trigger and Delayed-
Trigger modes, with different limitations on the spacing between

trigger events in each mode. Trigger events arriving too close to
the previous instance will be ignored by the digitizer.

X

In Post-Trigger mode, the minimum spacing between trigger
events is N+1

X

In Delayed-Trigger mode, the minimum spacing between
trigger events is (N+D)+1, where D is the number of the
delayed setting

Time

Operation start
Acquisition start

Trigger

Data

Acquisition stop
Data transfer to system begins

N samples

M samples

Trigger event occurs