Input[1..n]:attenuation – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual
Page 100

SCPI Command Reference
94
IQS-3150
Product-Specific Commands—Description
:INPut[1..n]:ATTenuation
Description
This command sets the absolute attenuation to a
specific value. The valid range of values depends
on the type of instrument and the current
wavelength. This value is used only when the
ATTenuation control mode is active.
In POWer mode, the device adjusts the
attenuation automatically to match the desired
output power. For this reason, changes made to
attenuation via the INPut[1..n]:ATT command are
not taken into account.
At *RST, the absolute attenuation value that will
be set depends on the instrument you have.
Syntax
:INPut[1..n]:ATTenuation<wsp><Attenuation[<
wsp>DB]>|MAXimum|MINimum|DEFault
Parameter(s)
Attenuation:
The program data syntax for <Attenuation> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is DB. The
<Attenuation> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
highest supported value.