Product-specific commands-description, Product-specific commands—description, Calibration[1..n]:zero – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual
Page 96

SCPI Command Reference
90
IQS-3150
Product-Specific Commands—Description
Product-Specific Commands—Description
:CALibration[1..n]:ZERO
Description
Returns the mechanism of the instrument to its
home position.
This command is an event and has no associated
*RST condition or query form.
Syntax
:CALibration[1..n]:ZERO
Parameter(s)
None
Example(s)
STAT:QUES:BIT9:COND? Returns 1 if the
instrument's mechanism should be returned to
its home position.
STAT? Must return READY
CAL:ZERO
STAT:OPER:BIT9:COND? Keep resending the
query as long as the operation is not complete
(returned value is not 0).
Notes
This command will take at least 15 seconds to
complete.
See Also
:SENSe[1..n]:CORRection:COLLect:ZERO
:STATus:OPERation:BIT<n>:CONDition?
:STATus:QUEStionable:BIT<n>:CONDition?