SCPI Command Reference

Variable Attenuator

139

Product-Specific Commands—Description

:OUTPut[1..n]:RPOWer?

Description

This query returns a value indicating either the

current or the minimum/maximum relative

power value.

At *RST, the value that will be set depends on the

instrument you have.

Syntax

:OUTPut[1..n]:RPOWer?[<wsp>MAXimum|MINi

mum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a <CHARACTER PROGRAM DATA>

element. The allowed <CHARACTER PROGRAM

DATA> elements for this parameter are:

MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's

lowest supported value.

MAXimum is used to retrieve the instrument's

highest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax

<RelativePower>

Response(s)

RelativePower:
The response data syntax for <RelativePower>

is defined as a <NR3 NUMERIC RESPONSE

DATA> element.

The <RelativePower> parameter corresponds to

a valid relative power value.