Input[1..n]:attenuation – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual
Page 102

SCPI Command Reference
96
IQS-3150
Product-Specific Commands—Description
:INPut[1..n]:ATTenuation?
Description
This query returns a value indicating either the
current or the minimum/maximum absolute
attenuation value.
At *RST, the absolute attenuation value that will
be set depends on the instrument you have.
Syntax
:INPut[1..n]:ATTenuation?[<wsp>MAXimum|MI
Nimum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<Attenuation>