Input[1..n]:attenuation – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 101

Advertising
background image

SCPI Command Reference

Variable Attenuator

95

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the <Attenuation> parameter.

The <Attenuation> parameter corresponds to a

valid attenuation value. You can use the

INPut[1..n]:ATTenuation? MAX and

INPut[1..n]:ATTenuation? MIN queries to

determine a valid attenuation range for the

current wavelength.

Example(s)

INP:WAV 1310 NM

CONT:MODE ATT

INP:ATT 5 DB

STAT:OPER:BIT8:COND? Keep sending the query

as long as attenuation is not reached. When

attenuation is reached, the returned value is 0.

INP:ATT MIN

STAT:OPER:BIT8:COND? Keep sending the query

as long as attenuation is not reached. When

attenuation is reached, the returned value is 0.

See Also

:INPut[1..n]:ATTenuation?

:INPut[1..n]:RATTenuation

:OUTPut[1..n]:POWer

:STATus:OPERation:BIT<n>:CONDition?

:INPut[1..n]:ATTenuation