Status:operation:bit<n>:condition – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 153

Advertising
background image

SCPI Command Reference

Variable Attenuator

147

Product-Specific Commands—Description

:STATus:OPERation:BIT<n>:CONDition?

Description

This query returns the state of a specific bit in the

OPERation register set. The <n>, ("bit <n>")

indicates for which bit the information must be

retrieved in the :STATus:OPERation status

register. The <n> value must be a number from

8 to 12.

At *RST, the value that will be set depends on the

instrument you have.

Syntax

:STATus:OPERation:BIT<n>:CONDition?

Parameter(s)

None

Response Syntax

<OperationCondition>

Response(s)

OperationCondition:
The response data syntax for

<OperationCondition> is defined as a <NR1

NUMERIC RESPONSE DATA> element.

The <OperationCondition> represents the

current operation condition of the instrument.

The meaning of the response depends on the

value returned for bit <n>.

Bit <8>: When the returned value is 1, the

instrument is currently adjusting the attenuation

to reach a new set point. When the returned

value is 0, the new set point is reached and the

attenuation is stable.

Advertising