Status:operation:bit<n>:condition – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual
Page 153

SCPI Command Reference
Variable Attenuator
147
Product-Specific Commands—Description
:STATus:OPERation:BIT<n>:CONDition?
Description
This query returns the state of a specific bit in the
OPERation register set. The <n>, ("bit <n>")
indicates for which bit the information must be
retrieved in the :STATus:OPERation status
register. The <n> value must be a number from
8 to 12.
At *RST, the value that will be set depends on the
instrument you have.
Syntax
:STATus:OPERation:BIT<n>:CONDition?
Parameter(s)
None
Response Syntax
<OperationCondition>
Response(s)
OperationCondition:
The response data syntax for
<OperationCondition> is defined as a <NR1
NUMERIC RESPONSE DATA> element.
The <OperationCondition> represents the
current operation condition of the instrument.
The meaning of the response depends on the
value returned for bit <n>.
Bit <8>: When the returned value is 1, the
instrument is currently adjusting the attenuation
to reach a new set point. When the returned
value is 0, the new set point is reached and the
attenuation is stable.