To freeze the first stable reading on the LCD
press

AutoEnd while the instrument is in

measurement mode.

The ”

HOLD” tag will be displayed blinking on

the LCD until the reading is stabilized.

When the reading is stable, the “

HOLD” tag

stops blinking and the reading is frozen on
the LCD.

Press

AutoEnd again to return to normal measurement mode.

Note: • Pressing Range the instrument will skip to the displayed

range, without leaving AutoEnd mode. The

Log key also

holds AutoEnd mode.

• Pressing

2nd then Setup, GLP or RCL, the instrument

leaves AutoEnd mode and performs the selected function.

The EC reading is affected by temperature and changes in ion concentration.
To measure only one variable, conductivity measurements are normally
compensated to a fixed temperature. This reference temperature is
configured in SETUP parameter rEF. See page 18.
Three options of temperature compensation are available:
Automatic (Atc): The probe has a built-in temperature sensor; the
value of the temperature is used to automatically compensate the EC/TDS
measurement to a reference temperature. This is the default option.
Manual (Mtc): The temperature value, shown on the lower LCD, can be
manually set by the user with the up and down arrow keys to the
solution temperature. The “

ºC” or “ºF” symbol blinks when this option is

active. The instrument compensates the conductivity measured at this
temperature to the reference temperature.
No Compensation (NoTC): The temperature reading shown on the
lower LCD is not taken into account. The reading displayed on the upper
LCD is the absolute EC or TDS value without temperature compensation
applied. The measurement changes due to temperature and ion
concentration. The second “

ºC” or “ºF” symbol blinks when this option is

active.