Hanna Instruments HI 98360 User Manual

Page 13

Advertising
background image

13

Notes: • If the uncalibrated reading is too far from the expected

value, the “

WRONG” tag will blink. Calibration can not be

confirmed. In this case check if the correct standard has been
used or clean the probe by following the Probe Maintenance
(see page 31).

If the meter is in

MTC or ATC mode

and the standard temperature is
out of the 0.0 - 60.0 ºC interval,

WRONG” “ºC” tags and the

temperature will be displayed
blinking.

For best results choose an EC standard value close to the

sample to be measured.

In order to minimize any EMC interference, use plastic or

glass beakers.

It is possible to set the cell constant value directly in according

with the specifications of the electrode indicated by the
manufactures, without following the calibration procedure.
To set the cell constant, enter SETUP mode and select “

CELL”

(see SETUP for details, page 18).

When the reading is stable, “

READY” tag

is displayed and “

CFM” along with “~”

tag starts blinking on the LCD, asking for
confirmation.

Press

CFM (or 2nd then CFM) to confirm

calibration.
The instrument stores the calibration value
and returns to measurement mode.