Hanna Instruments HI 98360 User Manual
Page 26

26
• the calibration time (12 chars)
• standard unit (always 00 - 2 chars)
• Reference Temperature with and decimal point
(4 chars)
• Temperature Compensation mode (2 chars)
• 00 - no temperature compensation
• 01 - automatic temperature compensation
• 02 - manual temperature compensation
• TC coeficient with sign and decimal point (5 chars)
PAR
Requests the setup parameters setting.
The answer string contains:
• Instrument ID (4 chars)
• SETUP information (2 chars): 8 bit hexadecimal
encoding.
• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• Auto-off/Light time (3 chars) (in minutes)
• Auto power off time (3 chars) (in minutes)
• Cell constant, with sign and decimal point (6 chars)
• TDS factor, with sign and decimal point (5 chars)
• TC coef, with sign and decimal point (5 chars)
• Reference Temperature, with sign and decimal point
(5 chars)
• Temperature Compensation mode (1 char)
NSLx
Requests the number of logged samples (4 chars).
The command parameter (1 char):
• E - request for EC range
• N - request for NaCl range
• T - request for TDS range
LODExxx
Requests the xxx
th
EC record logged data.
LODNxxx Requests the xxx
th
NaCl record logged data.
LODTxxx
Requests the xxx
th
TDS record logged data.
LODEALL Requests all EC Log on demand.
LODNALL Requests all NaCl Log on demand.
LODTALL Requests all TDS Log on demand.