Hanna Instruments HI 98360 User Manual
Page 18

18
S E T U P
S E T U P
S E T U P
S E T U P
S E T U P
Setup mode allows viewing and modifying the following parameters:
• Cell constant (CELL)
• TDS factor (FAct)
• Temperature coefficient (tc)
• Reference temperature (rEF)
• Current Time (hour & minute)
• Current Date (month, day & year)
• Beep Status (bEEP)
• Instrument ID (InSId)
• Auto-off backlight timer (LIGH)
• Auto power off timer (AOFF)
• Temperature Unit
To enter SETUP mode, press
2nd then Setup while the instrument is in
measurement mode.
Select the desired setup parameter using the
ARROW keys.
Press
CAL if you want to change the item
value. “
CFM” tag and the selected item (e.g.
hour, in setting up the correct time) will start
blinking.
Press the
ARROW keys to change the
displayed value.
If there is another item to be set (e.g.
minutes), press
Range. The other item will
start blinking.
Press the
ARROW keys to change the
displayed value.
Press
CFM (or 2nd then CFM) to confirm or
CAL to escape without saving.
Press the
ARROW keys to select the next/previous parameter.
Press
Setup to exit SETUP menu at any time.
The following table (see page 19) lists the SETUP parameters, their valid
values range and the factory settings (default). Use
to move through
parameters in this order.