Hanna Instruments HI 98360 User Manual
Page 27

27
The answer string for each record contains the number of records if ALL
data is requested - 3 chars
• The logged mode (2 chars):
• 06 - EC range
• 07 - TDS range
• 08 - NaCl range
• Reading status (1 char): R, O, U
• Calculated reading, with sign and decimal point
(7 chars)
• Temperature reading, with sign and two decimal
points (8 chars)
• EC reading status (1 char): R, O, U - for TDS,
NaCl range
• Unit of the reading - for TDS, NaCl range (1 char)
• The EC reading, with sign and decimal point - for
TDS, NaCl range (7 chars)
• The logged time,
yymmddhhmmss (12 chars)
• The calibration offset, with sign and decimal point
(7 chars)
• The calibration slope, with sign and decimal point
(7 chars) (Kcell for EC and TDS range and the slope
for NaCl).
• Temperature probe presence (1 char)
• Temperature compensation coefficient (5 chars)
• Temperature reference value (5 chars)
• Temperature compensation mode (1 char)
Notes: • “Err8” is sent if the instrument is not in measurement mode.
• “Err6” is sent if the requested range is not available.
• “Err4” is sent if the requested set parameter is not available.
• “Err3” is sent if the Log on demand is empty.
• “Err9” is sent if the battery power is less than 30%.
• Invalid commands will be ignored.