Chapter 6: Maintaining Your TXE

Product Service 1-800-522-7658

23

7. Carefully remove the old bag stub using the single glove near the access

port. Reattach the retaining strap on the perimeter of the new bag at the
enclosure access port. Draw the old bag stub into the single glove and turn
the glove inside out. Double seal this glove and cut off the glove with the
contaminated bag stub inside.

8. Carefully install the new HEPA filter and uniformly tighten the two filter

clamp bolts until the HEPA filter gasket is compressed 50%.

9. Carefully fold and roll the new bag in four-inch pleats and push the bag

into the access port. Reinstall the retaining strap and bungee cords. Turn
the enclosure ON to draw a vacuum on the bag. Reinstall the front panel.

























Figure 6-1

Bag-In/Bag-Out HEPA Filter Changing Procedure Diagram & Filter

Leak Test Diagram

Old HEPA

Filter

New HEPA

Filter

Old HEPA filter

sealed in old bag

Create two seals and
cut here after sealing

Bag stub from

old bag

Old bag stub

sealed in glove

of new bag.

Create two

seals and cut

here after

sealing

New HEPA

filter shown in

new bag

Filter gasket

up as shown